Thank you for purchasing the YSI Pro2030, an instrument from the YSI Professional
Series
product family. The Pro2030 measures dissolved oxygen, conductivity and
temperature in water. The Pro2030 features an impact resistant and waterproof
(IP-67) case, a rugged MS-8 (military-spec) cable connector, backlit display,
user-selectable sensor options, 50 data set memory, internal barometer and a
rubber over-mold case.
The Pro2030 provides valuable instructions and prompts near the bottom of
the display that will guide you through operation and use. However, reading the
entire manual is recommended for a better understanding of the instrument’s
features.
G E T T I N G S T A R T E D
InItIal InsPectIon
Carefully unpack the instrument and accessories and inspect for damage.
Compare received parts with items on the packing list. If any parts or materials
are damaged or missing, contact YSI Customer Service at 800-897-4151 (+1 937
767-7241) or the authorized YSI distributor from whom the instrument was
purchased.
Battery InstallatIon
The instrument requires 2 alkaline C-cell batteries. Under normal conditions,
the average battery life is 425 hours at room temperature without using the back
light. A battery symbol
will blink in the lower, left corner of the display
to indicate low batteries when approximately 1 hour of battery life remains.
To install or replace the batteries:
Turn the instrument off and flip over to view the battery cover on the
1.
back.
Unscrew the four captive battery cover screws.
2.
Remove the battery cover and remove the old batteries if necessary.
3.
Install the new batteries, ensuring correct polarity alignment (figure
4.
1).
The Pro2030 cannot communicate to a PC via a Pro Series
communications saddle. Connecting the Pro2030 to a communication
saddle may cause erratic instrument behavior.