YSI Pro2030 User Manual
Page 20

32
33
Thread the cap membrane onto the sensor. It is normal for
a small amount of electrolyte to overflow.
Replace the sensor guard.
Polarographic Sensors – Model # 605203
The cap membrane and KCl (potassium chloride) electrolyte solution should be
changed every 2-4 weeks during regular use. In addition, the electrolyte solution
and membrane should be changed if (a) bubbles are visible under the membrane;
(b) significant deposits of dried electrolyte are visible on the membrane; and (c)
if the sensor shows unstable readings or other sensor-related symptoms.
During membrane changes, examine the gold cathode at the tip of the sensor and
the silver anode along the shaft of the sensor (figure 12). If either the silver anode
is black in color or the gold cathode is dull, the sensor may need resurfaced using
the 400 grit wet/dry sanding discs included in the membrane kit. Do not sand
the electrode every membrane change as this is not routine maintenance. In fact,
visually, the anode may appear tarnished and operate properly. YSI recommends
using the sanding disc if the sensor has difficulty stabilizing or calibrating after
a regular membrane change.
To clean and resurface the sensor, follow the instructions on the next page.
Gold Cathode
For correct sensor operation, the gold cathode (figure 12) must be textured
properly. It can become tarnished or plated with silver after extended use. Never
use chemicals or abrasives that have not been recommended or supplied by
YSI.
First dry the sensor tip completely with lens cleaning tissue. Wet a sanding disc
and place it face up in the palm of you hand. Next, with your free hand, hold
the sensor in a vertical position, tip down. Place the sensor tip directly down on
the sanding disc and twist it in a circular motion to sand the gold cathode. The
goal is to sand off any build-up and to lightly scratch the cathode to provide a
larger surface area for the electrolyte solution under the membrane. Usually, 3 to
4 twists of the sensor are sufficient to remove deposits and for the gold to appear
to have a matte finish. Rinse the sensor thoroughly with distilled or deionized
water and wipe the gold cathode with a wet paper towel before putting on a new
cap membrane. If the cathode remains tarnished, contact YSI Technical Support
or the YSI authorized dealer where you purchased the instrument.
Figure 12, DO sensor with membrane removed.
Silver Anode
After extended use, a thick layer of Silver Chloride (AgCl) builds up on the
silver anode (figure 12) reducing the sensitivity of the sensor. The anode must
be cleaned to remove this layer and restore proper performance. The cleaning
can be chemical and/or mechanical:
Mechanical cleaning: In order to sand the silver anode along the shaft of the
sensor, simply hold the sensor in a vertical position. Wet the sanding disc and
gently wrap it around the sensor and twist it a few times to lightly sand the anode
(the goal is to simply sand off any build-up without scratching or removing
layers of the anode itself). Usually, 3 to 4 twists of the sanding disc are sufficient
to remove deposits.
After completing the sanding procedure, repeatedly rinse the electrode with
distilled or deionized water and wipe with lens cleaning tissue to remove any
grit left by the sanding disc. Thoroughly rinse the entire tip of the sensor with
distilled or deionized water and install a new membrane.
IMPORTANT: Be sure to: (1) Use only the fine sanding discs provided
and (2) Sand as mentioned in the above procedures. Not adhering to
either of these instructions can damage the electrodes.
i