beautypg.com

Do units, Conductivity units (cond. units) – YSI Pro2030 User Manual

Page 11

background image

14

15

do unIts

Highlight Do Units and press enter to open a submenu that allows you to select

the dissolved oxygen units to be displayed on the Run screen. Highlight a unit

and press enter to enable or disable it. An enabled dissolved oxygen unit will

have an ‘X’ in the box next to it. Highlight the ESC-Exit box along the bottom

of the display and press enter to save any changes and to close the Do units

submenu.

There are three options for displaying dissolved oxygen:

mg/L will show Do readings in milligrams per liter on a scale from 0

to 50 mg/L.

ppm (parts per million) is equivalent to mg/L and will show the Do

reading on a scale from 0 to 50 ppm.

% will show Do readings in a % saturation from 0 to 500%. This value

will be expressed as %L when Do Local% is enabled.

mg/L and ppm cannot be enabled and therefore displayed at the same time.

Do% and mg/L or ppm can be enabled and displayed simultaneously.

conductIvIty unIts (cond. unIts)

Highlight Cond. Units (Conductivity Units) and press enter to open a submenu

that allows you to select the conductivity units to be displayed on the Run screen.

Highlight a unit and press enter to enable or disable it. An enabled conductivity

unit will have an ‘X’ in the box next to it. Highlight the ESC-Exit box along

the bottom of the display and press enter to save any changes and to close the

conductivity units submenu.

There are seven options for displaying conductivity. only two units can be

enabled at the same time:

CoND-mS/cm displays conductivity in milliSiemens per centimeter.

CoND-uS/cm displays conductivity in microSiemens per centimeter.

SPC-mS/cm displays Specific Conductance in milliSiemens per

centimeter. Specific Conductance is temperature compensated

conductivity.

SPC-uS/cm displays Specific Conductance in microSiemens per

centimeter. Specific Conductance is temperature compensated

conductivity.

Sal ppt displays salinity in parts per thousand. The salinity reading is

calculated from the instrument’s conductivity and temperature values

using algorithms found in Standard Methods for the Examination of

Water and Wastewater.

TDS g/L displays

Total Dissolved Solids in grams per liter. TDS is

calculated from conductivity and temperature using a user-selectable

TDS constant.

TDS mg/L displays

Total Dissolved Solids in milligrams per liter. TDS

is calculated from conductivity and temperature using a user-selectable

TDS constant.

Note: 1 milliSiemen = 1,000 microSiemens.

SPECIFIC CoNDUCTANCE

The conductivity of a sample is highly dependent on temperature, varying as

much as 3% for each change of one degree Celsius (temperature coefficient =

3%/°C). In addition, the temperature coefficient itself varies with the nature of

the ionic species present in the sample. Therefore, it is useful to compensate for

this temperature dependence in order to quickly compare conductivity readings

taken at different temperatures.

The Pro2030 can display non-temperature compensated conductivity as well

as temperature compensated Specific Conductance. If Specific Conductance is

selected, the Pro2030 uses the temperature and conductivity values associated

with each measurement to calculate a specific conductance value compensated

to a user selected reference temperature, see below. Additionally, the user can

select the temperature coefficient from 0% to 4%.

Using the Pro2030’s default reference temperature and temperature coefficient

(25 °C and 1.91%), the calculation is carried out as follows:

Specific Conductance (25°C) = Conductivity of sample

1 + 0.0191 * (T - 25)

T = Temperature of the sample in °C

sPecIfIc conductance reference

teMPerature (sPc ref. teMP.)

SPC Ref. Temp. (Specific Conductance Reference Temperature) is the reference

temperature used to calculate Specific Conductance. The reference temperature

range is 15 and 25 °C. The default value is 25 °C.

To change the reference temperature, highlight SPC Ref. Temp. and press enter

to open the submenu. With the reference temperature highlighted, press enter

to make the field adjustable. Next, use the up or down arrow key to increase

or decrease the value. Press enter to save the new reference temperature. Next,

highlight the ESC-Exit box and press enter to close the submenu.