beautypg.com

Specifications – YSI Pro2030 User Manual

Page 23

background image

38

39

Symptom

Possible Solution

Instrument will not calibrate the

Conductivity sensor; instrument

displays “Calibration over”,

“Calibration Under”, or “Unstable

Reading” during calibration.

Ensure the conductivity sensor is

1.

clean. Follow the cleaning procedures

in the Care, Maintenance and

Storage section of this manual.

Verify the calibration solution is above

2.

the two holes near the cable, see figure 8.

Verify the calibration solution

3.

is not expired or contaminated.

Try a new bottle of solution.

Ensure you are entering in the correct

4.

value for the solution according to the

measurement units. 1 mS = 1,000 uS.

Allow sufficient stabilization time

5.

for conductivity and temperature

AND wait at least 3 seconds before

confirming a calibration.

Contact YSI Tech Support.

6.

Conductivity readings

are inaccurate.

Ensure the conductivity sensor is

1.

clean. Follow the cleaning procedures

in the Care, Maintenance and

Storage section of this manual.

Verify the sample is above the two

2.

holes near the cable, see figure 8.

Verify calibration.

3.

Verify temperature readings

4.

are accurate.

Verify the correct units are setup in the

5.

System Setup menu, i.e. uS vs mS and

Conductivity vs. Specific Conductance.

Contact YSI Tech Support.

6.

Conductivity values display

over or Undr on Run screen.

Ensure the conductivity sensor is

1.

clean. Follow the cleaning procedures

in the Care, Maintenance and

Storage section of this manual.

Verify the sample is above the two

2.

holes near the cable, see figure 8

Verify calibration.

3.

Verify temperature readings

4.

are accurate.

Sample conductivity is outside

5.

the measurement range of the

instrument, i.e. 0-200 mS.

Contact YSI Tech Support.

6.

S P E C I F I C A T I o N S

These specifications represent typical performance and are subject to change

without notice. For the latest product specification information, please visit

YSI’s website at www.ysi.com or contact YSI Tech Support.

Parameter

Range

Resolution

Accuracy

Temperature -5 to 55°C*

0.1°C

± 0.3°C

Dissolved

oxygen

0 to 200% air

saturation

1% or 0.1%,

user selectable

± 2% of the reading or

± 2% air saturation,

whichever is greater

200 to 500%

air saturation

1% or 0.1%,

user selectable

± 6% of the reading

0 to 20 mg/L

0.1 or 0.01

mg/L, user

selectable

±2% of the reading

or ± 0.2 mg/L,

whichever is greater

20 to 50 mg/L

0.1 or 0.01

mg/L, user

selectable

±6% of the reading

Conductivity 0-500 uS/cm

0-5 mS/cm

0-50 mS/cm

0-200 mS/cm

(auto ranging)

0.0001 to

0.1 mS/cm;

0.1 to 0 uS/

cm (range

dependent)

Instrument only: ± 0.5%

of the reading or 1 uS/

cm, whichever is greater.

Instrument with 1

or 4 meter cables:

± 1.0% of the

reading or 1 uS/cm,

whichever is greater.

Instrument with 10, 20,

or 30 meter cables: ± 2.0%

of the reading or 1 uS/

cm, whichever is greater.

Salinity

0 to 70 ppt

0.1 ppt

± 1.0% of the

reading or ± 0.1 ppt,

whichever is greater.

Total

Dissolved

Solids (TDS)

0 to 100 g/L.

TDS Constant

range: 0.3 to

1.00 (0.65

default)

0.0001 to 0.1

g/L (range

dependent)

Dependent on accuracy of

temperature, conductivity

and TDS Constant.

Barometer

500.0 to 800.0

mmHg**

0.1 mmHg

±5 mmHg within 15°C of

calibration temperature

* Automatic dissolved oxygen temperature compensation range is -0 to 45 ºC

**Available barometer units include: mmHg, inHg, mbars, psi, or KPa