4 setting table of the varion k (potassium sensor), Setting table of the varion k (potassium sensor) – YSI IQ S ENSOR N ET VARiON Plus 700 IQ User Manual
Page 28
Commissioning
VARiON
®Plus
700 IQ
3 - 16
ba76000e01
01/2012
3.5.4
Setting table of the VARiON K (potassium sensor)
Note
These sensor settings are only available if in the sensor settings for the
VARiON A the VARiON K setting was set to Active (see section 3.5.2).
The basic settings are taken over from the VARiON A sensor but can
be adjusted separately afterwards.
Menu item
Selection/values
Explanations
Measuring mode
K (mg/l)
mV
Citation form of the mass concentration or
voltage of the electrode.
Measuring range
(Measuring mode: K (mg/l))
1 ... 1000 mg/l
Fixed range
Measuring range
(Measuring mode: mV)
-2000 ... 2000 mV
Fixed range
IQ-LabLink
Not active
(not used)
Temperature mode
(only with
Measuring mode: K (mg/l))
°C
°F
Unit of the measured temperature value
(Celsius, Fahrenheit).
Temp. adjustment
(only with
Measuring mode: K (mg/l))
-1.5 °C ... +1.5 °C
The temperature compensation function
enables the temperature sensor to be
balanced against a reference temperature
measurement (displacement of the zero
point by ±1.5 °C).
Notes:
Due to the thermal capacity of the
sensor, it is necessary to place it in a
container with at least 2 liters of water.
Leave the sensor in this container for at
least 15 minutes while stirring
occasionally, then carry out the
adjustment.
If the temperature difference of the
water and sensor is > 10°C, leave the
sensor in the container for at least one
hour while stirring occasionally.
Conc. offset
-10 mg/l ... +10 mg/l
Constant value, which is added to the
measuered value.
Program level
0 ... 2
Program extension (0 = Off = standard set-
ting).
Refer to section 4.2.4 for details about the
program extension.