YSI IQ S ENSOR N ET VARiON Plus 700 IQ User Manual
Page 24
Commissioning
VARiON
®Plus
700 IQ
3 - 12
ba76000e01
01/2012
Potassium compens.
(only with setting
VARiON K: hide)
automatic / manual
1 ... 1000 mg/l
automatic
(with a potassium electrode mounted):
When the potassium electrode is
mounted, the potassium compensation
takes places automatically only. The value
of the potassium concentration measured
at the time the setting table was opened is
displayed in the next line (Potassium
conc.).
manual
(without any potassium electrode
mounted):
After determining the potassium content of
the test solution enter the determined
potassium content manually in the next
line (Potassium conc.).
The measured value is accordingly
corrected by the entered potassium
content.
Note: Detailed information on the subject
of potassium compensation is given in the
YSI primer, I
ON
SELECTIVE
MEASUREMENT
IN
ONLINE
ANALYSIS
.
Potassium conc.
(only with setting
VARiON K: hide)
1 ... 1000 mg/l
With Potassium compens.: automatic:
Display of the potassium measured value.
With Potassium compens.: manual:
Entry of the K value
IQ-LabLink
Not active
(not used)
Temperature mode
(only with Measuring mode:
NH4-N or NH4)
°C
°F
Unit of the measured temperature value
(Celsius, Fahrenheit).
Menu item
Selection/values
Explanations