8 lists, 1 explanation of the messages, 1 error messages – YSI IQ S ENSOR N ET MIQ/C6 User Manual
Page 37: 2 informative messages, Lists -1, Explanation of the messages -1 8.1.1, Error messages -1, Informative messages -1, 8lists
MIQ/C6
Lists
8 - 1
ba76033e01
01/2012
8
Lists
8.1
Explanation of the messages
In this chapter you will find a list with all the message codes and
corresponding message texts that may occur in the log book of the
IQ S
ENSOR
N
ET
system for the MIQ/C6 output module.
Note
Information about
Contents and structure of the log book and
Structure of the message code
can be found in the L
OG
BOOK
chapter of the IQ S
ENSOR
N
ET
system
operating manual.
Note
All message codes of the MIQ/C6 output module end with the number,
"421".
8.1.1
Error messages
8.1.2
Informative messages
The MIQ/C6 output module does not send informative messages.
Message code
Message text
EA4421
Current output range undercut
* Check process
* Check settings and, if necessary, change them
EA5421
Current output range exceeded
* Check process
* Check settings and, if necessary, change them
EI3421
Burden resistor too large (> 500 Ohm)
or current loop interrupted
* Check burden, terminal connections and connection lines