1 switching on the maintenance condition, 2 switching off the maintenance condition, Switching on the maintenance condition -18 – YSI IQ S ENSOR N ET MIQ/C6 User Manual
Page 30: Switching off the maintenance condition -18

Settings
MIQ/C6
4 - 18
ba76033e01
01/2012
4.7.1
Switching on the maintenance condition
4.7.2
Switching off the maintenance condition
1
Call up the measured value display with
m
.
2
Select the sensor you want to switch on the maintenance
condition for with
d
. The measured values of the sensor do
not flash.
3
Open the Display/Options menu with
g
.
4
Select the Switch maintenance condition on/off menu item with
d
and confirm with
g
. A window that informs you about the
maintenance condition opens up.
5
Confirm Continue with
g
.
The selected sensor is in the maintenance condition. Linked
outputs are frozen.
6
Call up the measured value display with
m
.
The measured values of the sensor flash.
1
Call up the measured value display with
m
.
2
Select the sensor you want to switch off the maintenance
condition for with
d
. The measured values of the sensor
flash.
3
Open the Display/Options menu with
g
.
4
Select the Switch maintenance condition on/off menu item with
d
and confirm with
g
. A window that informs you about the
maintenance condition opens up.
5
Confirm Continue with
g
.
The maintenance condition of the selected sensor is switched
off. Linked outputs are released.
6
Call up the measured value display with
m
.
The measured values of the sensor do not flash.