1 recorder – YSI IQ S ENSOR N ET MIQ/C6 User Manual
Page 18
Settings
MIQ/C6
4 - 6
ba76033e01
01/2012
4.4.1
Recorder
Function
In the Recorder application, the measured values of the linked sensor
are set up as current intensity at the current output. The output of
measured values is laid down in the Recorder type, Start value and End
value settings.
Settings
8
Carry out the settings for the current output with
d
and
g
.
9
Highlight and confirm Save and quit with
d
and
g
.
The new settings are stored.
Setting
Selection/Values
Explanation
Recorder type
0 to 20 mA or
4 to 20 mA
Start value
within the
measuring range
(sensor-
dependent)
Minimum spacing:
5 % of the measuring
range
End value
Measured variable
Main variable
Adjoining variable
Main variable designates
the actual measured
variable of the sensor
(e.g. pH, oxygen, etc.).
Adjoining variable
designates an additional
measured variable (e.g.
temperature).
Attenuation
0 ... 40 mA/s
Speed of change of the
output current (mA/s) in
the case of an erratic
change of the input
signal.
I -> OFL/UFL
Error
Current values outside
the range between Start
value and End value are
considerd as errors. The
current output reacts as
specified under Behavior
at error (see below).
Limitation
The current at the output
is limited to the Start
value or End value.