2 info messages, Info messages -4 – YSI NiCaVis 701 IQ NI User Manual
Page 58
Indexes
NiCaVis 70x IQ NI
9 - 4
ba76132e01
03/2013
9.1.2 Info messages
Message code
Message text
IAC541
Zero adjustment was successfully carried out
* Carry out sensor check
* If required, carry out new user calibration
IC3541
The factory zero adjustment was activated. Make sure the sensor op
-
erates correctly.
IC4541
The last valid user zero adjustment was activated. Make sure the sen
-
sor operates correctly.
IC5541
Invalid user zero adjustment was replaced by last valid user calibration.
Caution! Wrong measured values possible. Carry out a new successful
zero adjustment to make sure the sensor operates correctly.
IC6541
The invalid user zero adjustment was replaced by the factory zero
adjustment. Caution! Wrong measured values possible. Check
whether the sensor operates correctly or carry out a new successful
user zero adjustment.