Technical data
8.1
Measurement characteristics
Measuring principle
Spectrophotometric absorption measurement; integrated micropro-
cessor electronics, shielded 2-wire connection for
power and data transmission.
Wavelength range
200 - 390 nm
Measuring gap (optical
layer thickness)
Nitrite/nitrate
measurement
1 mm gap width
NiCaVis 701 IQ NI
1 mm
NiCaVis 705 IQ NI
5 mm
Measuring
location
Measuring mode
Measuring range
Resolution
Inlet
NO2
0.0 ... 120.0 mg/L
0.1 mg/L
NO2-N
0.00 ... 30.00 mg/l 0.01 mg/L
NO3
0.0 ... 300.0 mg/L
0.1 mg/L
NO3-N
0.00 ... 60.00 mg/L 0.01 mg/L
Activation
NO2
0.0 ... 120.0 mg/L
0.1 mg/L
NO2-N
0.00 ... 30.00 mg/l 0.01 mg/L
NO3
0.0 ... 300.0 mg/L
0.1 mg/L
NO3-N
0.00 ... 60.00 mg/L 0.01 mg/L
Outlet
NO2
0.0 ... 300.0 mg/L
0.1 mg/L
NO2-N
0.00 ... 75.00 mg/L 0.01 mg/L
NO3
0.0 ... 750.0 mg/L
0.1 mg/L
NO3-N
0.0 ... 150.0 mg/l
0.1 mg/L