4 settings for virtual sensors, Settings for virtual sensors -12 – YSI NiCaVis 701 IQ NI User Manual
Page 22
Commissioning
NiCaVis 70x IQ NI
3 - 12
ba76132e01
03/2013
3.4.4 Settings for virtual sensors
Carrying out settings
Using , switch from the measured value display to the main menu
of the settings. Then navigate to the setting menu (setting table) of the
sensor. The procedure is described in detail in your IQ S
ENSOR
N
ET
system operating manual.
Default values are marked in bold.
Menu item
Settings
Explanations
Measuring mode
(NitraVis 70x IQ NI)
NO3-N
NO3
NO2-N
NO2
The measured parameter is displayed
in the selected citation form.
Measuring mode
(CarboVis 70x IQ NI)
COD spectr., total
COD spectr., dissolv
TOC spectr., correl.
BOD spectr., correl.
DOC spectr., correl.
SAC 254, total
SAC 254, dissolved
UVT 254
The measured parameter is displayed
in the selected citation form.
Measuring location
NiCaVis 701 IQ NI
:
Activation
Inlet
Outlet
NiCaVis 705 IQ NI
:
Outlet
(permanently set)
Measurement location or application of
the sensor.
The possible measurement loca-
tions are displayed depending on the
currently set measuring mode.
The sample matrix on which the cal-
culation of the measured value is
based changes with the location of
the sensor.
Measuring range
-
Display of the measuring range
Cal - # raw value 1
Value pairs of the user calibration (see section 4.3.2).
Note: Leave those values unchanged during the initial commission-
ing.
Cal - ref. value 1
Cal - # raw value 2
Cal - ref. value 2