YSI NiCaVis 701 IQ NI User Manual
Page 31
NiCaVis 70x IQ NI
Measuring / Operation
4 - 7
ba76132e01
03/2013
Default values are marked in bold.
Calibration evaluation
The entered calibration data are checked for plausibility by the system.
The check can have the following results:
Information on the contents and structure of the log book and how you
can call it up is given in the L
OG
BOOK
chapter of the IQ S
ENSOR
N
ET
system operating manual.
The user calibration is not recorded in the calibration history. You can,
however, view the current value pairs in the
Settings of sensors and
diff. sensors
menu at any time.
Menu item
Settings
Explanations
Cal - # raw value 1
0.00 ... 5000.00
Raw value of the first value pair
(lower concentration).
For single-point calibration, enter 0.
Cal - ref. value 1
0.00 ... 5000.00
Reference value of the first value pair
(lower concentration).
For single-point calibration, enter 0.
Cal - # raw value 2
0.00 ... 10.00 ... 5000.00 Raw value of the second value pair
(higher concentration) or raw value of the
single-point calibration.
Cal - ref. value 2
0.00 ... 10.00 ... 5000.00 Reference value of the second value pair
(higher concentration) or reference value
of the single-point calibration.
Display
Meaning/actions
Measured value dis-
play
Successful new valid calibration value pairs
were entered for the sensor.
"----"
Calibration value pairs not accepted. Sen-
sor is blocked for measurement. A log book
message is generated.
Causes:
Ascending order not observed.
At least one value is outside the measur-
ing range.