Maintenance and cleaning
5.1
Maintenance
The UV NiCaVis 70x IQ NI sensor operates maintenance-free.
5.2
Sensor cleaning
5.2.1 Cleaning agents and accessories
Cleaning agents
To clean the sensor, use the following cleaning agents only:
* For the final cleaning prior to a
Sensor check
or
Zero adjustment
, only
use cleaning agents with purity grade "for analysis". Finally, thoroughly
rinse the sensor with ultrapure water.
Accessories
Sponge
Plastic-bristled brushes of different sizes.
Interdental brushes are suitable for the cleaning of the compressed
air channels.
Flocked cleaning cards (see accessories)
Lens tissues
Wash bottle
Note
Never use any pressure washer, steam blaster, abrasives or metal
brushes to clean the sensor. Risk of damaging the sensor.
Contamination
Cleaning agents
Water-soluble substances
– Primary cleaning: Tapwater,
warm if possible
(30 - 50 °C)
– Precision cleaning: House-
hold washing-up liquid + ultra-
pure water.
Lime deposits
5 % hydrochloric acid *
Grease deposits
Mixture of isopropanol and water
(approx. 70 % isopropanol) *