3.4.3 Settings for the main sensor
Measuring cycle
A measuring cycle consists of the cleaning procedure, the adjustment
time for the measuring system and the determination of the measured
value. The following graphic demonstrates the relevant settings:
Fig. 3-6 Measuring cycle of the UV sensor
Carrying out settings
Using , switch from the measured value display to the main menu
of the settings. Then navigate to the setting menu (setting table) of the
sensor. The procedure is described in detail in your IQ S
ENSOR
N
ET
system operating manual.
A change of the following settings will take some time (up to a few min-
utes) until it can be seen in the measured value display:
Measuring mode
Measuring location
Number of N-sensors
Number of C-sensors
Further changes can only be made when the last changes are visible
in the measured value display.
Default values are marked in bold.
Time
Adjustment time
Measurement
Measuring
cycle
Measuring interval
Cleaning
Menu item
Settings
Explanations
Measuring mode
NO3-N
NO3
NO2-N
NO2
The measured parameter is displayed
in the selected citation form.
When changing the
Measuring mode
setting, always check the
Measuring
location
setting.