8 technical data, 1 measurement characteristics, Technical data -1 – YSI CarboVis 705 IQ TS User Manual
Page 49: 1 measurement characteristics -1, Section 8.1 m, Easurement, Characteristics, Ee section, See chapter 8 t, Echnical
CarboVis 70x IQ TS
Technical data
8 - 1
ba75950e02
11/2012
8
Technical data
8.1
Measurement characteristics
Measuring principle
Spectrophotometric absorption measurement; integrated micropro-
cessor electronics, shielded 2-wire connection for
power and data transmission.
Wavelength range
200 - 720 nm
Measuring gap (optical
layer thickness)
Carbon measurement,
1 mm gap width
CarboVis 701 IQ TS
1 mm
CarboVis 705 IQ TS
5 mm
Measuring
location
Measuring mode
Measuring range Resolution
Inlet
COD spectr., total
0 ... 20000 mg/L
1 mg/L
COD spectr., dissolv
0 ... 12500 mg/L
1 mg/L
TOC spectr., correl.
0 ... 20000 mg/L
1 mg/L
DOC spectr., correl.
0 ... 12500 mg/L
1 mg/L
BOD spectr., correl.
0 ... 8000 mg/L
1 mg/L
SAC 254, total
0 ... 5000 1/m
1 1/m
SAC 254, dissolved
0 ... 3000 1/m
1 1/m
UVT 254
0.0 ... 100.0 %
0.1 %
Activation
COD spectr., dissolv
0 ... 12500 mg/L
1 mg/L
DOC spectr., correl.
0 ... 12500 mg/L
1 mg/L
SAC 254, total
0 ... 5000 1/m
1 1/m
SAC 254, dissolved
0 ... 3000 1/m
1 1/m
UVT 254
0.0 ... 100.0 %
0.1 %