2 measuring operation, 2 measuring operation -2 – YSI CarboVis 705 IQ TS User Manual
Page 26
Measuring / Operation
CarboVis 70x IQ TS
4 - 2
ba75950e02
11/2012
4.2
Measuring operation
CAUTION
Never look into the measuring window or put objects into the measur-
ing window during operation! The emitted UV radiation can damage the
eyes!
In the case of authorized use, inadvertent eye contact with the light
beam is not possible.
Measured value display
Fig. 4-2 CarboVis 70x IQ TS
in the measured value display with main sensor, sec
-
ondary sensor and virtual sensor
Factors affecting the
measured value
The following factors can adversely affect the measurement quality:
The measuring gap is not sufficiently filled and flowed through by the
test sample
Foreign bodies and air bubbles are in the measuring gap
Extraordinary changes (interferences) of the sample matrix
Contaminated measurement windows due to difficult waste water
matrix
Damaged measuring windows
Lacking quality of the zero adjustment, e. g. due to contaminated
ultrapure water
The user calibration is not representative (or erroneous)
Raw value,
identifiable by #
Measured value