beautypg.com

4 settings for virtual sensors, Settings for virtual sensors -12 – YSI CarboVis 705 IQ TS User Manual

Page 22

background image

Commissioning

CarboVis 70x IQ TS

3 - 12

ba75950e02

11/2012

3.4.4 Settings for virtual sensors

Carrying out settings

Using , switch from the measured value display to the main menu

of the settings. Then navigate to the setting menu (setting table) of the

sensor. The procedure is described in detail in your IQ S

ENSOR

N

ET

system operating manual.
Default values are marked in bold.

Menu item

Settings

Explanations

Measuring mode

COD spectr., total

COD spectr., dissolv

TOC spectr., correl.

BOD spectr., correl.

DOC spectr., correl.

SAC 254, total

SAC 254, dissolved

UVT 254

The measured parameter is displayed

in the selected citation form.

Measuring location

Outlet

Inlet

Activation

Measurement location or application of

the sensor.

The possible measurement loca-

tions are displayed depending on the

currently set measuring mode.

The sample matrix on which the cal-

culation of the measured value is

based changes with the location of

the sensor.

The measurement location

Activation

can only be selected for

Measuring

mode

COD spectr., dissolv

.

Measuring range

-

Display of the measuring range

(see chapter 8 T

ECHNICAL

DATA

).

Cal - # raw value 1

Value pairs of the user calibration (see section 4.3.2).

Note: Leave those values unchanged during the initial commission-

ing.

Cal - ref. value 1
Cal - # raw value 2
Cal - ref. value 2
Main sensor

Information that there are further sen-

sor overlapping settings in the setting

menu of the relevant main sensor

(

Meas. interval

,

Cleaning duration

,

Sig

-

nal smoothing

etc.).

This manual is related to the following products: