YSI CarboVis 705 IQ TS User Manual
Page 20
Commissioning
CarboVis 70x IQ TS
3 - 10
ba75950e02
11/2012
Menu item
Settings
Explanations
Measuring mode
COD spectr., total
COD spectr., dissolv
TOC spectr., correl.
BOD spectr., correl.
DOC spectr., correl.
SAC 254, total
SAC 254, dissolved
UVT 254
The measured parameter is displayed
in the selected citation form.
When changing the
Measuring mode
setting, always check the
Measuring
location
setting.
If the previously set measurement loca-
tion is not available for the selected
measuring mode, the measurement
location is automatically changed..
Measuring location
Outlet
Inlet
Activation
Measurement location or application of
the sensor.
The possible measurement loca-
tions are displayed depending on the
currently set measuring mode.
The sample matrix on which the cal-
culation of the measured value is
based changes with the location of
the sensor.
The measurement location
Activation
can only be selected for
Measuring
mode
COD spectr., dissolv
.
Measuring range
-
Display of the measuring range
Cal - # raw value 1
Value pairs of the user calibration (see section 4.3.2).
Note: Leave those values unchanged during the initial commission-
ing.
Cal - ref. value 1
Cal - # raw value 2
Cal - ref. value 2
Zero adjustment
Factory
User
Setting of which zero adjustment is
used in the sensor.
Factory zero adjustment or
Zero adjustment carried out last
Signal smoothing
On
Off
Switching on/off the signal smoothing.
For more details, see
Response time
t90
.