What to do if...
IFL 70x IQ
6 - 2
ba76129e01
10/2012
The measured value is
not within the expected
range
Cause
Remedy
The
Water depth is not set correctly
(e.g., the bottom echo or multiple
echoes between the bottom of the
basin and the surface of the water
are interpreted as measured value
echoes.)
Set the Water depth and
Immersion depth correctly
In the Evaluation range there are
permanently installed fixtures that
continually generate interfering
echoes
– Select a measurement loca-
tion without permanent inter-
fering echoes.
– If necessary, limit the Evalu-
ation range so that any per-
manent interfering echoes
are outside the Evaluation
range.
In the Evaluation range there are
moving fixtures (scrapers) that tem-
porarily generate echoes.
– Limit the Evaluation range
so that the water depth of
the scraper is not in the
Evaluation range.
– Set the Establishing time
correctly
– Set Follow echo to Yes.
The sensor is cyclically moved out
of the water by the scraper.
– Set the Establishing time
correctly
– Set Follow echo to Yes.
- 3100 (50 pages)
- 3200 (86 pages)
- 5000 (70 pages)
- 5200 (148 pages)
- 5200A (121 pages)
- 5200A (128 pages)
- 5400 (127 pages)
- 5239 Dissolved Oxygen Probe (23 pages)
- 55 (26 pages)
- 550A (18 pages)
- 556 MPS - Quick Start Guide (4 pages)
- 58 (40 pages)
- 5905 (22 pages)
- 6600V2 - User Manual (379 pages)
- 60 (33 pages)
- 63 (46 pages)
- 650 MDS (72 pages)
- 9300 Photometer (11 pages)
- 9500 Photometer (143 pages)
- 9500 Photometer (18 pages)
- CR 3200 Thermoreactor (42 pages)
- CR 4200 Thermoreactor (42 pages)
- CR2200 Thermoreactor (36 pages)
- DO200 (40 pages)
- EC300 (40 pages)
- EC300A (40 pages)
- EC30A (12 pages)
- IQ SensorNet 2020 XT Terminal (196 pages)
- IQ SensorNet FDO Sensor (48 pages)
- IQ SensorNet MIQ/A-MOD Module (42 pages)
- IQ SensorNet MIQ/A-PR Module (36 pages)
- IQ SensorNet MIQ/IC2 Module (30 pages)
- IQ SensorNet MIQ/IF232 Module (18 pages)
- IQ SensorNet MIQ/JB Module (18 pages)
- IQ SensorNet MIQ/JBR Module (20 pages)
- IQ SensorNet MIQ/PS Module (22 pages)
- IQ SensorNet MIQ/R6 Module (56 pages)
- IQ SensorNet NitraLyt Sensor (64 pages)
- IQ SensorNet SensoLyt Sensor 700 IQ (SW) (46 pages)
- IQ SensorNet Sensors Encoded Fieldbus Data (12 pages)
- IQ SensorNet TetraCon Sensor 700 IQ (SW) (32 pages)
- IQ SensorNet ViSolid Sensor 700 IQ (60 pages)
- MultiLab 4010-1 (108 pages)
- MultiLab 4010-2 (138 pages)