Commissioning
3 - 13
ba76129e01
10/2012
3.5.2 Display/Extras menu
Carrying out settings
Using <>, select the IFL 70x IQ sensor in the measured value dis-
play. Using
menu. Open the Sensor settings menu.
Setting
Selection/values
Explanation
X-axis (depth)
Complete
The complete height from the ultrasonic
transducer to the bottom is displayed.
Evaluation range
The complete Evaluation range is dis-
played.
Zoom range
Begin
End
The display on the screen is limited to the
section set here.
Y-axis (intensity)
Auto
The strongest echo is displayed with the
complete intensity.
Zoom range
Begin
End
The display on the screen is limited to the
section set here.
Displayed profile
This setting only effects the display of the
echo profile.
The current measured value is still deter-
mined with the filter settings.
Unfiltered
Displays the echo profile without any fil-
ters.
Filtered
Displays the echo profile with all filters.
Apply
Closes the Display/Extras menu.
Scraper test
(only with IFL 700 IQ )
The wiper moves once (function test).