6 what to do if, What to do if -1, 6what to do if – YSI IFL 700 IQ User Manual
Page 31

IFL 70x IQ
What to do if...
6 - 1
ba76129e01
10/2012
6
What to do if...
Mechanical damage to
the sensor
Display "----"
(no valid measured
value)
Cause
Remedy
– Return the sensor
Cause
Remedy
Sensor is permanently in the air
Immerse the sensor in water
(see section 3.3.2)
There are too many air bubbles in
the water or on the ultrasonic trans-
ducer
Select a measurement location
free of air bubbles
Sensor is dirty
– Clean the sensor and/or its
environment
– Check the function of the
wiper (see section 3.5)
Thread algae floating in front of the
sensor
Remove the thread algae from
the sensor or its environment
In the selected Evaluation range,
no echo is available that meets all
settings
– Check whether there is a
sludge blanket in the
selected Evaluation range
– Check whether the settings
are suitable for the applica-
tion.
Defective sensor
Contact the service department
- 3100 (50 pages)
- 3200 (86 pages)
- 5000 (70 pages)
- 5200 (148 pages)
- 5200A (128 pages)
- 5200A (121 pages)
- 5400 (127 pages)
- 5239 Dissolved Oxygen Probe (23 pages)
- 55 (26 pages)
- 550A (18 pages)
- 556 MPS - Quick Start Guide (4 pages)
- 58 (40 pages)
- 5905 (22 pages)
- 6600V2 - User Manual (379 pages)
- 60 (33 pages)
- 63 (46 pages)
- 650 MDS (72 pages)
- 9300 Photometer (11 pages)
- 9500 Photometer (143 pages)
- 9500 Photometer (18 pages)
- CR 3200 Thermoreactor (42 pages)
- CR 4200 Thermoreactor (42 pages)
- CR2200 Thermoreactor (36 pages)
- DO200 (40 pages)
- EC300 (40 pages)
- EC300A (40 pages)
- EC30A (12 pages)
- IQ SensorNet 2020 XT Terminal (196 pages)
- IQ SensorNet FDO Sensor (48 pages)
- IQ SensorNet MIQ/A-MOD Module (42 pages)
- IQ SensorNet MIQ/A-PR Module (36 pages)
- IQ SensorNet MIQ/IC2 Module (30 pages)
- IQ SensorNet MIQ/IF232 Module (18 pages)
- IQ SensorNet MIQ/JB Module (18 pages)
- IQ SensorNet MIQ/JBR Module (20 pages)
- IQ SensorNet MIQ/PS Module (22 pages)
- IQ SensorNet MIQ/R6 Module (56 pages)
- IQ SensorNet NitraLyt Sensor (64 pages)
- IQ SensorNet SensoLyt Sensor 700 IQ (SW) (46 pages)
- IQ SensorNet Sensors Encoded Fieldbus Data (12 pages)
- IQ SensorNet TetraCon Sensor 700 IQ (SW) (32 pages)
- IQ SensorNet ViSolid Sensor 700 IQ (60 pages)
- MultiLab 4010-1 (108 pages)
- MultiLab 4010-2 (138 pages)