7 technical data, 1 measurement characteristics, 2 application characteristics – YSI IFL 700 IQ User Manual
Page 35: Technical data -1, Measurement characteristics -1, Application characteristics -1, Ee chapter 7 t, Echnical, Data, See chapter 7 t

IFL 70x IQ
Technical data
7 - 1
ba76129e01
10/2012
7
Technical data
7.1
Measurement characteristics
Measuring principle
Ultrasound echo measurement
Measuring ranges and
resolution
Conversion to sludge level depth (from the surface of the water) or
sludge level height (from the bottom of the basin)
7.2
Application characteristics
Allowed
temperature range
Allowed pH range of the
measuring medium
4 ... 12
Pressure resistance
Sensor with connected SACIQ (SW) sensor connection cable:
Type of protection
Sensor with connected SACIQ (SW) sensor connection cable:
IP X8; 0.3 bar (3*10
5
Pa)
Immersion depth
min. 5 cm; max. 3 m
Flow speed
max. 3 m/s
Operating position
Vertical with ultrasonic transducer pointing downward
Measured
parameter
Measuring ranges
Resolution
Accuracy
Distance
0.4 ... 15 m from
ultrasonic trans-
ducer surface
0.01 m
0.1 m
Measuring medium
0 °C ... + 50 °C (32 ... 122°F)
Storage/transport
- 5 °C ... + 50 °C (23 ... 122 °F)
Max. allowed overpressure:
3*10
5
Pa (0.3 bar)