TEMPERATURE CORRECTION

This data is for typical applications of the YSI sondes.

A. Potassium Chloride

**

(KCl)

Concentration: 1 mole/liter

Concentration: 1 x 10

-1

mole/liter

°C

mS/cm

%/

°C (to 25°C)

°C

mS/cm

%/

°C (to 25°C)

0 65.10

1.67

0 7.13

1.78

5 73.89

1.70

5 8.22

1.80

10 82.97

1.72

10 9.34

1.83

15 92.33

1.75

15 10.48

1.85

20 101.97

1.77

20 11.65

1.88

25 111.90

1.80

25 12.86

1.90

30

14.10

1.93

35

15.38

1.96

37.5

16.04

1.98

40

16.70

1.99

45

18.05

2.02

50

19.43

2.04

Concentration: 1 x 10

-2

mole/liter

Concentration: 1 x 10

-3

mole/liter

°C

mS/cm

%/

°C (to 25°C)

°C

mS/cm

%/

°C (to 25°C)

0 0.773

1.81

0 0.080

1.84

5 0.892

1.84

5 0.092

1.88

10 1.015

1.87

10 0.105

1.92

15 1.143

1.90

15 0.119

1.96

20 1.275

1.93

20 0.133

1.99

25 1.412

1.96

25 0.147

2.02

30 1.553

1.99

30 0.162

2.05

35 1.697

2.02

35 0.178

2.07

37.5 1.771

2.03

37.5 0.186

2.08

40 1.845

2.05

40 0.194

2.09

45 1.997

2.07

45 0.210

2.11

50 2.151

2.09

50 0.226

2.13