3 changing from autoranging to manual ranging, Model 30 – YSI 30 User Manual
Page 21
Advanced Setup
Section 6
YSI, Incorporated
Model 30/30M
17
4. Use the UP ARROW or DOWN ARROW key to change the value to the desired new
reference temperature (any value between 15
o
C and 25
o
C is acceptable).
15.0C
22.7
CAL
New reference temperature
5. Press the ENTER key. The word “SAVE” will flash across the display for a second to indicate
that your change has been accepted. The instrument will automatically return to normal
operation mode.
6.3 CHANGING FROM AUTORANGING TO MANUAL RANGING
If your application is easier to perform using a manual range which you select, the YSI Model 30
and Model 30M allow you to turn off the default autoranging feature. Follow the instructions for
your model as shown below.
Model
30
While you are making conductivity or specific conductance measurements with the Model 30,
simply press and release the ENTER key to select manual ranging.
1014
µS
24.8
°C
CAL
Each additional press of the ENTER key will cycle the Model 30 to a different manual range until
you return again to autoranging. Five pushes of the ENTER key will cycle the Model 30 through the
four manual ranges and return the instrument to autoranging.
NOTE: You may see an error message in some manual ranges if the manual range selected is not
adequate for the sample you are measuring.