What to do if...

handylab pH/LF 12

68

ba75370e03

11/2009

No stable mea-

sured value

Obviously incor-

rect measured val-

ues

Buffer solutions

– Incorrect buffer solutions

– Change calibration procedure

– Buffer solutions too old

– Use only once.

Note the shelf life

– Buffer solutions depleted

– Change solutions

Cause

Remedy

pH electrode:

– Diaphragm contaminated

– Clean diaphragm

– Membrane contaminated

– Clean membrane

Test sample:

– pH value not stable

– Measure with air excluded if

necessary

– Temperature not stable

– Adjust temperature if

necessary

Electrode + test sample:

– Conductivity too low

– Use suitable electrode

– Temperature too high

– Use suitable electrode

– Organic liquids

– Use suitable electrode

Cause

Remedy

pH electrode:

– pH electrode unsuitable

– Use suitable electrode

– Temperature difference

between buffer and test
sample too high

– Adjust temperature of buffers

or test samples

– Measurement procedure not

suitable

– Follow special procedure