beautypg.com

6 setting the temperature compensation, Setting the temperature compensation – Xylem handylab pH LF 12 User Manual

Page 42

background image

Operation

handylab pH/LF 12

42

ba75370e03

11/2009

4.4.6

Setting the temperature compensation

The calculation of the temperature compensation is based on the preset ref-
erence temperature, Tref 20 or Tref 25 (see section 4.8 C

ONFIGURATION

).

You can select one of the following temperature compensations:

z Nonlinear temperature compensation (nLF)

according to EN 27 888

z Linear temperature compensation (Lin) with adjustable coefficient in

the range 0.001 ... 3.000 %

/

K

z No temperature compensation (----)

Note

ips

Select the following temperature compensations given in the table according
to the respective test sample:

Application tips

Test sample

Temperature compensation

Display in-
dicator

Natural water (ground
water, surface water
and drinking water)

nLF
according to DIN 38404
EN 27 888

Ultrapure water

nLF
according to DIN 38404
EN 27 888

Other aqueous solu-
tions

Set linear temperature coeffi-
cient
0.001 ... 3.000 %/K

Salinity
(seawater)

Automatically nLF according
to IOT

Sal,