4 conductivity/salinity, 1 general information, Conductivity/salinity 4.4.1 – Xylem handylab pH LF 12 User Manual
Page 35: General information
handylab pH/LF 12
Operation
35
ba75370e03
11/2009
4.4
Conductivity/Salinity
4.4.1
General information
Caution
When connecting an earthed PC/printer, measurements cannot be per-
formed in earthed media as incorrect values would result.
The RS232 interface is not galvanically isolated.
Preparatory
activities
Perform the following preparatory activities when you want to measure:
Temperature probe
Only carry out measurements with a temperature probe. The temperature
probe is shown on the display by TP.
Temperature
compensation
Selectable linear or nonlinear temperature compensation, can be switched
off (see section 4.8 C
Reference
temperature, Tref
The reference temperature (Tref) can be switched between 20 °C and 25 °C.
It appears on the display as Tref20 or Tref25. To switch over the reference
temperature, see S
1
Connect a conductivity measuring cell to the measuring instru-
ment. The instrument automatically switches over to conductiv-
ity measurement (status display
ӕ
or SAL).
If the conductivity measuring cell is already connected, press
the
ӕ
or SAL ap-
pears.
2
Check the selected cell constant (see below) or calibrate the
measuring instrument with the measuring cell (see
section 4.4.5).
3
Using <
▲ > <▼>, toggle between the measuring modes, con-
ductivity (
ӕ
in
μS/cm) or salinity (SAL).