4.5
Calibration intervals (Int 3, Int 5)
For each measured variable, a time interval is stored. When it has expired,
you will be reminded to calibrate. After a calibration interval has expired, the
probe symbol of the relevant measured variable flashes. It is still possible to
measure. By calibrating the relevant probe, the function is reset and the in-
terval starts anew.
The following calibration intervals are set in the factory:
Setting the
calibration interval
You can change each of these intervals (1 ... 999 days):
Measured
parameter
Designation
Default setting
pH/ORP
Int 3
7 days (d 7)
Conductivity
Int 5
180 days (d 180)
1
Switch off the measuring instrument.
2
Press the
3
Press the
The display test appears briefly on the display. After this, the
measuring instrument automatically switches over to configu-
ration.
4
Press the
the required measured variable pH or
appears. Example:
5
Set the required interval (in days) until the next calibration with
<
▲> <▼>.
6
Confirm with
7
Switch to the measuring mode with