Specifi cations – Xylem HandyLab 700 User Manual
Page 36
36
Specifi cations
Memosens pH input (ISFET)
M8 socket, 4 pins, for Memosens lab cable
Display ranges
1)
-2,000 … +16,000 pH
-2000 … +2000 mV
-50 … +250 °C
Memosens Redox
M8 socket, 4 pins, for Memosens lab cable
Display ranges
1)
-2000 … +2000 mV
-50 … +250 °C
Sensor standardization
*)
ORP calibration (zero adjustment)
Permissible calibration range ΔmV (offset) -700 … +700 mV
pH/mV (analog)
Anschluss
pH socket, DIN 19 262
(13/4 mm)
pH range
–2 … 16
Decimal places *
)
2 or 3
Input resistance
1 x 10
12
Ω (0 … 35 °C)
Input current
1 x 10
-12
A (bei RT, Verdopplung alle 10 K)
Measuring cycle
approx 1s
Measurement error
2,3,4)
< 0.01 pH, TC < 0.001
pH/K
mV range
-1300 … +1300
Measuring cycle
approx 1s
Measurement error
2,3,4)
< 0,1 % meas. val. + 0,3 mV, TC < 0,03 mV/K
Temperature
2 x 4 mm dia. for integrated or separate temperature detector
Measuring ranges
NTC 30 kΩ
–20 … +120 °C
Pt 1000
–40 … +250 °C
Measuring cycle
approx 1s
Measurement error
2,3,4)
< 0.2 K (Tamb = 23 °C); TC < 25 ppm/K
pH-calibration
AutoCal buffer sets
*)
-01-
(Default)
NIST standard
1.679/4.006/6.865/9.180
-02-
NIST technical
1.68/4.00/7.00/10.01/12.46
-03-
SI Analytics technical
2.00/4.00/7.00/10.00
-04-
Knick Calimat
2.00/4.00/7.00/9.00/12.00
-05-
Mettler Toledo
2.00/4.01/7.00/9.21
-06-
HACH
4.01/7.00/10.01/12.00
-07-
Ciba (94)
2.06/4.00/7.00/10.00
-08-
Hamilton
2.00/4.01/7.00/10.01/12.00
-09-
Reagecon
2.00/4.00/7.00/9.00/12.00
-10-
DIN 19267
1.09/4.65/6.79/9.23/12.75
Operating modes
*)
AutoCal
Calibration with automatic buffer recognition
MANUAL
Manual calibration with entry of individual buffer
values
DATA INPUT
Data entry of zero and slope
Permissible calibration
range
Zero point
pH 6 … 8
With ISFET:
–750 … +750 mV Operating point (asymmetry)
Slope
approx. 74 … 104 %
Calibration timer
*)
Interval 1 … 99 days, can be switched off