Specifi cations – Xylem HandyLab 700 User Manual

Page 36

background image

36

Specifi cations

Memosens pH input (ISFET)

M8 socket, 4 pins, for Memosens lab cable
Display ranges

1)

-2,000 … +16,000 pH
-2000 … +2000 mV
-50 … +250 °C

Memosens Redox

M8 socket, 4 pins, for Memosens lab cable
Display ranges

1)

-2000 … +2000 mV
-50 … +250 °C

Sensor standardization

*)

ORP calibration (zero adjustment)

Permissible calibration range ΔmV (offset) -700 … +700 mV

pH/mV (analog)

Anschluss

pH socket, DIN 19 262
(13/4 mm)

pH range

–2 … 16

Decimal places *

)

2 or 3

Input resistance

1 x 10

12

Ω (0 … 35 °C)

Input current

1 x 10

-12

A (bei RT, Verdopplung alle 10 K)

Measuring cycle

approx 1s

Measurement error

2,3,4)

< 0.01 pH, TC < 0.001
pH/K

mV range

-1300 … +1300

Measuring cycle

approx 1s

Measurement error

2,3,4)

< 0,1 % meas. val. + 0,3 mV, TC < 0,03 mV/K

Temperature

2 x 4 mm dia. for integrated or separate temperature detector
Measuring ranges

NTC 30 kΩ

–20 … +120 °C

Pt 1000

–40 … +250 °C

Measuring cycle

approx 1s

Measurement error

2,3,4)

< 0.2 K (Tamb = 23 °C); TC < 25 ppm/K

pH-calibration

AutoCal buffer sets

*)

-01-
(Default)

NIST standard

1.679/4.006/6.865/9.180

-02-

NIST technical

1.68/4.00/7.00/10.01/12.46

-03-

SI Analytics technical

2.00/4.00/7.00/10.00

-04-

Knick Calimat

2.00/4.00/7.00/9.00/12.00

-05-

Mettler Toledo

2.00/4.01/7.00/9.21

-06-

HACH

4.01/7.00/10.01/12.00

-07-

Ciba (94)

2.06/4.00/7.00/10.00

-08-

Hamilton

2.00/4.01/7.00/10.01/12.00

-09-

Reagecon

2.00/4.00/7.00/9.00/12.00

-10-

DIN 19267

1.09/4.65/6.79/9.23/12.75

Operating modes

*)

AutoCal

Calibration with automatic buffer recognition

MANUAL

Manual calibration with entry of individual buffer
values

DATA INPUT

Data entry of zero and slope

Permissible calibration
range

Zero point

pH 6 … 8

With ISFET:

–750 … +750 mV Operating point (asymmetry)

Slope

approx. 74 … 104 %

Calibration timer

*)

Interval 1 … 99 days, can be switched off