Calibrating, Manual calibration – Xylem HandyLab 700 User Manual

Page 31

background image

31

Calibrating

Measurement

The number of calibration points has been
selected in the confi guration menu.

pH display blinks

Use

to set the temperature-corrected

pH value taken from the buffer table.

mV display blinks

Depending on the number of calibration
points, the procedure described above for
CAL 1/2/3 is repeated.

mV value blinks until calibration is completed, then successive display of:

Then the meter switches to measuring mode.

Note:

To abort calibration, you can press meas at any time. This will be con-

fi rmed by the display message “CAL ABORTED”. Exception: When you have
selected “CAL POINTS 1-2-3” and the fi rst calibration step has been comple-
ted, the calibration process cannot be stopped any more.

MANUAL Calibration

( Manual calibration)
The calibration method is selected in the confi guration menu.