Confi guring, Ph confi guration – Xylem HandyLab 700 User Manual
Page 28
28
pH Confi guration
Prior to measurement, a confi guration should be performed
to match the connected sensor and the desired measurement
performance. Furthermore, you can select the suitable calibrati-
on method.
The following table gives you an overview. Factory settings are
shown in bold print.
Measurement
“Setup” display
set
Select using arrow keys, confi rm by pressing set.
Display 1
pH x.xx
/ pH x.xxx / mV /
(°C for analog pH only)
Display 2
OFF
/ date + time / date / time
CAL Timer
OFF
/ 1 ... 99 days
CAL
AutoCal
/Manual/DATA INPUT/
(ISFET-Zero)
/FREE CAL
CAL POINTS
1 / 2 / 3 / 1-2-3 (for AutoCal, Manual, FREE CAL)
BUFFER SET
(AutoCal,
FREE CAL)
-01-
(De-
fault)
NIST standard
1.679/4.006/6.865/9.180
-02-
NIST technical
1.68/4.00/7.00/10.01/12.46
-03-
SI Analytics technical
2.00/4.00/7.00/10.00
-04-
Knick Calimat
2.00/4.00/7.00/9.00/12.00
-05-
Mettler Toledo
2.00/4.01/7.00/9.21
-06-
HACH
4.01/7.00/10.01/12.00
-07-
Ciba (94)
2.06/4.00/7.00/10.00
-08-
Hamilton
2.00/4.01/7.00/10.01/12.00
-09-
Reagecon
2.00/4.00/7.00/9.00/12.00
-10-
DIN 19267
1.09/4.65/6.79/9.23/12.75
Auto OFF
OFF
/ 0.1h / 1h / 6h / 12h
Temp Unit
°C
/ °F
Time Format
24h
/ 12h
Date Format
dd.mm.yy
/ mm.dd.yy
Default
NO
/ YES (reset to factory settings)
This icon prompts you to select a menu item using the arrow keys –
the selection is confi rmed by pressing set.
Confi guring