28

pH Confi guration

Prior to measurement, a confi guration should be performed
to match the connected sensor and the desired measurement
performance. Furthermore, you can select the suitable calibrati-
on method.
The following table gives you an overview. Factory settings are
shown in bold print.

Measurement

“Setup” display

set

Select using arrow keys, confi rm by pressing set.

Display 1

pH x.xx

/ pH x.xxx / mV /

(°C for analog pH only)

Display 2

OFF

/ date + time / date / time

CAL Timer

OFF

/ 1 ... 99 days

CAL

AutoCal

/Manual/DATA INPUT/

(ISFET-Zero)

/FREE CAL

CAL POINTS

1 / 2 / 3 / 1-2-3 (for AutoCal, Manual, FREE CAL)

BUFFER SET

(AutoCal,

FREE CAL)

-01-

(De-

fault)

NIST standard

1.679/4.006/6.865/9.180

-02-

NIST technical

1.68/4.00/7.00/10.01/12.46

-03-

SI Analytics technical

2.00/4.00/7.00/10.00

-04-

Knick Calimat

2.00/4.00/7.00/9.00/12.00

-05-

Mettler Toledo

2.00/4.01/7.00/9.21

-06-

HACH

4.01/7.00/10.01/12.00

-07-

Ciba (94)

2.06/4.00/7.00/10.00

-08-

Hamilton

2.00/4.01/7.00/10.01/12.00

-09-

Reagecon

2.00/4.00/7.00/9.00/12.00

-10-

DIN 19267

1.09/4.65/6.79/9.23/12.75

Auto OFF

OFF

/ 0.1h / 1h / 6h / 12h

Temp Unit

°C

/ °F

Time Format

24h

/ 12h

Date Format

dd.mm.yy

/ mm.dd.yy

Default

NO

/ YES (reset to factory settings)

This icon prompts you to select a menu item using the arrow keys –
the selection is confi rmed by pressing set.

Confi guring