Error codes and device messages, Error messages – Xylem HandyLab 700 User Manual

Page 35

background image

35

Error Codes and Device Messages

Error Messages

The following error messages can be shown in the display.

Message

Cause Remedy

blinks

Battery empty

Replace batteries

ERROR 1

pH value out of range

Check whether the measurement
conditions correspond to the ad-
justed measuring range.

ERROR 2

ORP value out of range

ERROR 3

Temperature value out of
range

ERROR 4

Sensor zero point
too high/low

Thoroughly rinse the sensor and
recalibrate. If this does not help,
replace the sensor.

ERROR 5

Sensor slope too high/low

ERROR 8

Calibration error:
Identical buffers

Use a buffer solution with a diffe-
rent nominal value before starting
the next calibration step.

ERROR 9

Calibration error:
Buffer unknown

Make sure that you use the same
buffer set as confi gured.

ERROR
10

Cal media interchanged

Repeat calibration.

ERROR
11

Measured value unstable
Drift too high

Leave the sensor in the liquid until
the temperature is stable. If this
does not help, replace the sensor.

ERROR
14

Time and date invalid

Set time and date

ERROR
18

Confi guration invalid

Restart,reset to factory settings
(Setup: DEFAULT YES), confi gure
and calibrate.
If this does not help, send in the
device for repair.

ERROR
19

Factory settings error

Device defective, send it in.

ERROR
21

Sensor error (Memosens)

Connect operational Memosens
sensor.