4.5

Data transmission

You can use the following options to transmit data:

l

One of the following options:

– The

AutoStore

function (108) is used to periodically (

Int

1

storage interval) save measured values internally

and output them on the RS interface.

– The

data transmission interval

(Int 2)

function is used to

periodically output measured values to a RS interface
(see below).

l

The

Output data store

function (110) is used to output cal-

ibration data and stored measured values to the RS inter-
face.

l

The analog recorder output (116) is used to output mea-
sured values as voltages.

4.5.1 Data transmission interval (Int 2)

The interval for the data transmission (Int 2) is determined
by the time interval between automatic data transmissions.
After the time interval expires, the current data record is
transmitted to the RS interface.

Note
The setting of the interval (Int 2) only has an effect when the
storage interval (

AutoStore

function) is switched off.