2 automatic calibration (autocal), Automatic calibration (autocal) – Xylem Lab 860 User Manual
Page 28
![background image](/manuals/310557/28/background.png)
Operation
Lab 860
28
ba75554e05
01/2010
4.4.2
Automatic calibration (AutoCal)
Use any one to five buffer solutions of the selected buffer set.
Note
The steps 2, 6 and 10 are not necessary if you use a temperature
sensor.
1
Start the calibration with
The number of the specified buffer set is displayed (SET 1 ...
16).
The first buffer (BUFFER 1) of the buffer set (SET_x) is
requested.
2
If necessary, set the temperature of the buffer solution with
<
▲><▼>.
3
Immerse the pH electrode in the first buffer solution.
4
Start the measurement with
The SC display indicator flashes.
Depending on the setting, the display shows the nominal pH
value for the recognized buffer or the electrode voltage (mV).
As soon as a stable value is recognized, the next buffer
(BUFFER 2) is requested.
8
5
2
1
1
°C
TP
AutoCal
pH
SET
BUFFER