1 measuring the ph value, Measuring the ph value – Xylem Lab 860 User Manual
Page 20
Operation
Lab 860
20
ba75554e05
01/2010
4.3.1
Measuring the pH value
Stability control SC
(drift control)
The stability control function (drift control) checks the stability of the
measurement signal. The stability has a considerable effect on the
reproducibility of the measured value.
For identical measurement conditions, the following criteria apply:
Drift within 15 sec < 0.02 pH units.
Note
The current measurement with stability control can be terminated at
any time (accepting the current value) by pressing
1
Perform the preparatory activities according to page 19.
2
Immerse the pH electrode in the test sample.
3
Using
parameter pH is displayed.
8
4
2
2
9
9
6
pH
°C
TP
1
If necessary, call up the measured variable pH with
2
With
The SC function display indicator appears.
The current measured value is frozen (hold function).
3
Start measurement with stability control with
SC flashes until a stable measured value is reached.
This measured value is downloaded to the interface.
4
If necessary, start the next measurement with stability control
with
5
To terminate the stability control function: Press the
or
key.