Procedure - Manual Sampling:

1. Access the MAIN MENU by pressing the MENU key.
2. Advance the reverse shade cursor using the ARROW keys to highlight MANUAL SAMPLE.
3. Press the ENTER key to select the highlighted menu option.
4. The following displays appears:

MAIN MENU

AUTO SAMPLE

MANUAL SAMPLE

CALIBRATE


24.5 C 100 KPA



>>>

MANUAL RANGE

25%

10%

5%
1%

5. Advance the reverse shade cursor using the ARROW keys to highlight the desired RANGE.
6. Press the ENTER key to select the highlighted menu option.
7. The following display appears with the range selected and oxygen concentration of the sample gas:

3.3

%

MANUAL SAMPLING

10% RANGE

24.5 C 100 KPA



8. The display will not shift automatically. Instead, when the oxygen reading (actually the sensor’s signal

output) exceeds 110% of the upper limit of the current range an OVER RANGE warning will be displayed.

13.00

%

OVERRANGE

MANUAL SAMPLING

10% RANGE

24.5 C 100 KPA




Once the OVER RANGE warning appears the user must advance the transmitter to the next higher range via the

menu and keypad Press MENU, select MANUAL SAMPLING, press ENTER, select the appropriate MANUAL RANGE

and press ENTER again.