Advance the reverse shade cursor using the ARROW keys to highlight CALIBRATION.

Press the ENTER key to select the highlighted menu option.

The following displays appear:

MAIN MENU

AUTO SAMPLE

MANUAL SAMPLE

CALIBRATION

>>>

CALIBRATION

SPAN CALIBRATE

ZERO CALIBRATE

DEFAULT SPAN

DEFAULT ZERO

OUTPUT SPAN

OUTPUT ZERO

Advance the reverse shade cursor using the ARROW keys to highlight OUTPUT SPAN.

Press the ENTER key to select the highlighted menu option.

The following display appears:

100.0

OUTPUT SPAN OFFSET

PRESS UP OR DOWN

TO CHANGE VALUE

ENTER TO SAVE

MENU TO RETURN

Enter the calculated adjustment value, refer to example described above.

Note: Once the initial adjustment is made and checked at the PLC it may be necessary to fine tune the initial
adjustment by repeating. Any additional percent error must be added or subtracted from the initial
adjustment value

064.0

OUTPUT SPAN OFFSET

PRESS UP OR DOWN

TO CHANGE VALUE

ENTER TO SAVE

MENU TO RETURN

.

Press the ENTER key to advance the underline cursor right or press the MENU key to advance the
underline cursor left to reach to the desired digit of the adjustment OUTPUT SPAN OFFSET value.

Press the ARROW keys to enter the numerical value of each digit of the OUTPUT SPAN OFFSET value.

Repeat until the complete OUTPUT SPAN OFFSET value has been entered.

Save the adjustment value by pressing the ENTER key or abort by pressing the MENU key.

The system returns to the SAMPLING mode.