beautypg.com

Analytical Industries GPR-1500 A Series Trace PPM Oxygen Analyzer User Manual

Page 28

background image

Advanced Instruments, Inc

28

3. Press the ENTER key to select the highlighted menu option and the following displays appear:

MAIN MENU

AUTO SAMPLE

MANUAL SAMPLE

CALIBRATION

CONFIG ALARMS

BYPASS ALARMS

>>>

CALIBRATION

SPAN CALIBRATE

ZERO CALIBRATE

DEFAULT SPAN

DEFAULT ZERO

OUTPUT SPAN

OUTPUT ZERO

4. Advance the reverse shade cursor using the ARROW keys to highlight DEFAULT ZERO.

5. Press the ENTER key to select the highlighted menu option and the following display appears:

100.0

OUTPUT ZERO OFFSET

PRESS UP OR DOWN

TO CHANGE VALUE

ENTER TO SAVE

MENU TO RETURN

The default setting of 100 illustrates no adjustment to the analog output signal. Compute the adjustment
value as described in Appendix B or consult the factory. The true adjustment value must be determined
empirically by trial and error. Adjust the initial value to above 100 to increase the analog signal value or
decrease it below 100 to decrease the analog signal.

090.0

OUTPUT ZERO OFFSET

PRESS UP OR DOWN

TO CHANGE VALUE

ENTER TO SAVE

MENU TO RETURN

6. Press the ENTER key to advance the underline cursor right or press the MENU key to advance the

underline cursor left to reach to the desired digit of the OUTPUT ZERO OFFSET value.

7. Press the ARROW keys to enter the OUTPUT ZERO OFFSET value.

8. Repeat the OUTPUT ZERO OFFSET routine until the output is 4 mA .

Save the adjustment value by pressing the ENTER key or abort by pressing the MENU key.

After adjustment, the system returns to the SAMPLING mode.