Px-2 specifications, Single mode, Non-s2000 spectrometer strobe rate adjustments – Ocean Optics PX-2 User Manual
Page 6
PX-2 Installation and Operation Instructions
6
110-00000-000-01-201009
Non-S2000 Spectrometer Strobe Rate Adjustments
The table below shows the Continuous Strobe frequency (multiple strobe operation) for each type of
OOI spectrometer.
Spectrometer
Strobe Rate Adjustment
USB2000
Fixed at 122Hz
1
All other OOI Spectrometers
Software-controlled 220Hz-1Hz
2
1
Fixed frequency when the USB2000 is programmed with the PX2 firmware
2
The maximum frequency of the PX2; the spectrometer can actually run faster
Single Mode
In Single mode, a single lamp flash occurs during each integration period. You can modify the flash
rate by changing the integration period. Since the PX-2 has a maximum repetition rate of 220 Hz, the
minimum integration time allowed in Single mode is 5 milliseconds.
PX-2 Specifications
Specification
Value
Dimensions (mm)
153.4 x 104.9 x 40.9 mm
Weight
370 g
Power consumption
1.5 A @ 12 VDC
Power output
45 microjoules per pulse maximum
9.9 watts average power
220 Hz pulse rate maximum
Spectral range
220-750 nm
Pulse duration
5 microseconds (at 1/3 height of pulse)
Aperture
3 mm
Connector
SMA 905
Lifetime
10
9
pulses (estimated 230 days continuous operation)