beautypg.com

Px-2 specifications, Single mode, Non-s2000 spectrometer strobe rate adjustments – Ocean Optics PX-2 User Manual

Page 6

background image

PX-2 Installation and Operation Instructions

6

110-00000-000-01-201009

Non-S2000 Spectrometer Strobe Rate Adjustments

The table below shows the Continuous Strobe frequency (multiple strobe operation) for each type of
OOI spectrometer.

Spectrometer

Strobe Rate Adjustment

USB2000

Fixed at 122Hz

1

All other OOI Spectrometers

Software-controlled 220Hz-1Hz

2

1

Fixed frequency when the USB2000 is programmed with the PX2 firmware

2

The maximum frequency of the PX2; the spectrometer can actually run faster

Single Mode

In Single mode, a single lamp flash occurs during each integration period. You can modify the flash
rate by changing the integration period. Since the PX-2 has a maximum repetition rate of 220 Hz, the
minimum integration time allowed in Single mode is 5 milliseconds.

PX-2 Specifications

Specification

Value

Dimensions (mm)

153.4 x 104.9 x 40.9 mm

Weight

370 g

Power consumption

1.5 A @ 12 VDC

Power output

45 microjoules per pulse maximum
9.9 watts average power
220 Hz pulse rate maximum

Spectral range

220-750 nm

Pulse duration

5 microseconds (at 1/3 height of pulse)

Aperture

3 mm

Connector

SMA 905

Lifetime

10

9

pulses (estimated 230 days continuous operation)