If the power supply or lamp does not seem to functioning properly, check the following:

Issue

Probable Cause

Resolution

Power switches on, but no
LEDs light.

Line power not present

1. Check line voltage

2. Check if the lamp is

plugged in properly.

Fuse defective

Check fuse

Xenon lamp does not start

Lifetime of the lamp is
exhausted

Replace the lamp module

Lamp’s internal connection plug
is not closed properly

Open unit (see

Bulb Replacement

)

and close connector plug.