Operating instructions, Operating the xenon lamp, Starting main power – Ocean Optics HPX-2000-HP-DUV User Manual

Page 17: Starting the lamp, Chapter 3

background image

000-10000-081-02-201207b

9

Chapter 3

Operating Instructions

Operating the Xenon Lamp

The following sections provide instructions on operating the Xenon lamp in the HPX-2000-HP-DUV
Light Source. The HPX-2000-HP-DUV unit must be in a horizontal position for it to work.

Starting Main Power

Procedure

To apply main power to the unit,

1. Ensure that the proper voltage is set on the Voltage Selector Switch on the rear panel of the unit

(see

Rear Panel

).

2. Ensure that the Main Power Switch is OFF. Then, connect the power cord to the Terminal Input

on the rear panel of the unit.

3. Turn the Main Power Switch ON. The Power LED lights, indicating that the unit is receiving

power. The Xenon bulb remains off after power-on.

Starting the Lamp

Hold the Xenon ON/OFF switch down for 1 full second to illuminate the HPX-2000-HP-DUV lamp.

The lamp cannot be started when the Status LED is blinking. The Status LED blinks for about 5 seconds
after the main power is switched on (main power switch is located on rear of unit).

WARNING

Protective eyewear must be worn when using this equipment - Intense ultraviolet
radiation present. Never look directly into the light beam, as this can cause eye
damage.