Operating instructions, Operating the xenon lamp, Starting main power – Ocean Optics HPX-2000-HP-DUV User Manual
Page 17: Starting the lamp, Chapter 3
000-10000-081-02-201207b
9
Chapter 3
Operating Instructions
Operating the Xenon Lamp
The following sections provide instructions on operating the Xenon lamp in the HPX-2000-HP-DUV
Light Source. The HPX-2000-HP-DUV unit must be in a horizontal position for it to work.
Starting Main Power
►
Procedure
To apply main power to the unit,
1. Ensure that the proper voltage is set on the Voltage Selector Switch on the rear panel of the unit
(see
2. Ensure that the Main Power Switch is OFF. Then, connect the power cord to the Terminal Input
on the rear panel of the unit.
3. Turn the Main Power Switch ON. The Power LED lights, indicating that the unit is receiving
power. The Xenon bulb remains off after power-on.
Starting the Lamp
Hold the Xenon ON/OFF switch down for 1 full second to illuminate the HPX-2000-HP-DUV lamp.
The lamp cannot be started when the Status LED is blinking. The Status LED blinks for about 5 seconds
after the main power is switched on (main power switch is located on rear of unit).
WARNING
Protective eyewear must be worn when using this equipment - Intense ultraviolet
radiation present. Never look directly into the light beam, as this can cause eye
damage.