R-ls-1 pinout information – Ocean Optics LS-1 Series User Manual
Page 9
LS-1 Installation and Operation Instructions
009-00000-STD-01-201010
9
7. Slide the silver lamp tube out of the blue anodized bulb housing.
8. Detach the black wire from the silver bulb and dispose of the old bulb.
9. Plug the new bulb into the socket on the black wire.
10. Slide the silver bulb into the bulb housing on the front of the R-LS-1. Ensure that it is fully
inserted.
11. Tighten the setscrew on the bottom of the R-LS-1.
12. Tighten the four Phillips screws on the bottom of the R-LS-1
You have now properly replaced the bulb in the R-LS-1.
R-LS-1 Pinout Information
Pinouts for the H1 and H2 Headers of the R-LS-1 are the same as the H1 and H2 headers on the S2000
Spectrometer:
H1 Header Pins (Analog)
H2 Header Pins (Digital)
Pin
Description
Pin
Description
1
Analog Channel 0
D
N/C
2
Analog Channel 1
C
A/D Trigger
3
Analog Channel 2
B
Digital In 3 (D3)
4
Analog Channel 3
A
S1
5
Analog Channel 4
1
Ground
6
Ground
2
+5 VDC
7
Reserved
3
Phi A/D clock
8
Analog Channel 7
4
Phi Read Out Gate
9
Analog Channel 6
5
Reserved
10
Analog Channel 5
6
Temperature (optional)
7
Read Enable
8
S0
9
Strobe Single flash
10
Strobe Multiple Flash
11
Integration Clock
12
Master Clock
These are the same headers contained on an S2000 spectrometer. However, H1 is not used by the R-
LS-1. It is only required when the R-LS-1 is rack-mounted between spectrometers.