beautypg.com

External synchronization trigger – Ocean Optics Torus Operating Instructions User Manual

Page 24

background image

B: External Triggering

18

097-11101-000-02-201208

Timing Diagram for Variable Delay Mode

External Synchronization Trigger

In this mode of the Torus, the integration clock is essentially input into the spectrometer. The
synchronized trigger must have a frequency range of 5ms to 60s (shutter mode cannot be used in External
Synchronization Trigger mode). If no External Trigger Input Signal is applied, the spectrometer will not
take any spectra, and the software will appear “frozen” until an External Trigger Input Signal is detected.
With a continuous ExtTrigIn, the spectrometer will provide the required signals to the CCD as shown
below:

External Synchronization Trigger

The time between when ExtTrigIn goes HIGH and the beginning of the integration period is specified to
be between 8µs

SU_ST

<10µs.

The ExtTrigIn signal does not need to have an even duty cycle, but the time

between rising edges should remain constant for a given integration time.