7 Set up with the display and adjustment module

VEGAFLEX 86 • 4 … 20 mA/HART two-wire

44231-EN-130823

You can find detailed information on the proof test in the Safety

Manual (SIL).

Entering a 4-digit PIN protects the sensor data against unauthorized

access and unintentional modification. In this menu item, the PIN is

displayed or edited and changed. However, this menu item is only

available if adjustment is enabled in the menu "Lock/Release setup/

adjustment".

In delivery status, the PIN is "0000".

In this menu item, the internal clock of the sensor is adjusted.

With a reset, certain parameter adjustments carried out by the user

are reset.

The following reset functions are available:
Delivery status: Restoring the parameter settings at the time of ship-

ment from the factory incl. the order-specific settings. A created false

signal suppression, user-programmable linearization curve as well as

the measured value memory will be deleted.
Basic settings: Resetting of the parameter settings incl. special

parameters to the default values (presettings) of the respective instru-

ment. Any created false signal suppression or user-programmable lin-

earization curve as well as the measured value memory are deleted.
The following tables show the default values of the instrument. De-

pending on the instrument version or application, all menu items may

not be available or some may be differently assigned.
The menu items in bold are safety-relevant in terms of the functional

safety according to IEC 61508 (Edition 2) SIL.

Additional settings - PIN

Additional adjustments -

Date Time

Additional adjustments

- Reset