beautypg.com

VEGA VEGAPULS 61 (≥ 2.0.0 - ≥ 4.0.0) 4 … 20 mA_HART four-wire Approval according to LPR radio standard User Manual

Page 70

background image

-

Double chamber housing

l

1 x cable entry M20 x 1.5 (cable: ø 5 … 9 mm),
1 x blind stopper M20 x 1.5; 1 x blind stopper
M

16 x 1.5 or optional

4)

1 x plug M12 x 1 for

external indicating and adjustment unit

or:

l

1 x closing cap ½ NPT, 1 x blind stopper
½

NPT, 1 x blind stopper M16 x 1.5 or

optionally

5)

1 x plug M12 x 1 for external

indicating and adjustment unit

or:

l

1 x plug (depending on the version), 1 x blind
stopper M20 x 1.5; 1 x blind stopper M16 x 1.5
or optionally

6)

1 x plug M12 x 1 for external

indicating and adjustment unit

Spring-loaded terminals for wire cross-section

-

Massive wire, cord

0.2 … 2.5 mm² (AWG 24 … 14)

-

Stranded wire with end sleeve

0.2 … 1.5 mm² (AWG 24 … 16)

-

Two stranded wires with double end
sleeve, each

0.2 … 1.5 mm² (AWG 24 … 16)

Indicating and adjustment module

Voltage supply and data transmission

through the sensor

Indication

LC display in dot matrix

Measured value indication

-

Number of digits

5

-

Size of digits

W x H = 7 x 13 mm (sensors with software ≥ 4.0.0,
hardware ≥ 2.0.0)
W x H = 5 x 8 mm (sensors with software ≤ 3.99,
hardware < 2.0.0)

Adjustment elements

4 keys

Protection rating

-

unassembled

IP

20

-

mounted into the sensor without cover

IP

40

Materials

-

Housing

ABS

-

Inspection window

Polyester foil

Integrated clock

Date format

Day.Month.Year

Time format

12 h/24 h

4)

Only with Ex d version

5)

Only with Ex d version

6)

Only with Ex d version

70

VEGAPULS

61 • Approval according to LPR radio standard

11 Supplement

41714

-EN