VEGA VEGAPULS 61 (≥ 2.0.0 - ≥ 4.0.0) 4 … 20 mA_HART four-wire Approval according to LPR radio standard User Manual
Page 55
Code
Text mes-
sage
Cause
Rectification
F
125
Unpermissible
electronics
temperature
l Temperature of the elect-
ronics in the non-specified
section
l Check ambient temperatu-
re
l Isolate electronics
l Use instrument with higher
temperature range
F
260
Error in the
calibration
l Error in the calibration car-
ried out in the factory
l Error in the EEPROM
l Exchange of the electro-
nics
l Send instrument for repair
F
261
Error in the
configuration
l Error during setup
l False signal suppression
faulty
l Error when carrying out a
reset
l Repeat setup
l Repeat reset
F
264
Installation/S-
etup error
l Adjustment not within the
vessel height/measuring
range
l Max. measuring range of
the instrument not suffici-
ent
l Check or correct installa-
tion and/or parameter ad-
justment
l Use an instrument with
bigger measuring range
F
265
Measurement
function dis-
turbed
l Sensor does no longer
carry out a measurement
l Operating voltage too low
l Check operating voltage
l Carry out a reset
l Separate operating voltage
briefly
The following table shows the error codes and text messages in the
status message "Function check" and provides information on causes
as well as corrective measures.
Code
Text mes-
sage
Cause
Rectification
C
700
Simulation
l Simulation active
l Finish simulation
l Wait for the automatic end
after 60 mins.
The following table shows the error codes and text messages in the
status message "Out of specification" and provides information on
causes as well as corrective measures.
Function check
Out of specification
VEGAPULS
61 • Approval according to LPR radio standard
55
9 Diagnosis, Asset Management and service