VEGA VEGAPULS 61 (≥ 2.0.0 - ≥ 4.0.0) 4 … 20 mA_HART four-wire Approval according to LPR radio standard User Manual

Page 25

background image

2

3 ... 4 x h

max

90°

h

max

d

1

B

Fig. 19: Flow measurement with Khafagi-Venturi flume: d = Min. distance to
sensor; h

max.

= max. filling of the flume; B = tightest constriction in the flume

1

Position sensor

2

Venturi flume

In general, the following points must be observed:

l

Installation of the sensor at the inlet side

l

Installation in the centre of the flume and vertical to the liquid
surface

l

Distance to the Venturi flume

l

Min. distance of the sensor to max. storage level

Flow measurement with
Khafagi Venturi flume

VEGAPULS

61 • Approval according to LPR radio standard

25

4 Mounting

41714

-EN