11 Supplement

VEGAFLEX 82 • Modbus and Levelmaster protocol

41833-EN-130314

Ʋ Cable: ø 6 mm (0.236 in)

20 Nm (14.75 lbf ft)

Ʋ Cable: ø 11 mm (0.433 in) - PA coated 20 Nm (14.75 lbf ft)

Ʋ Rod: ø 16 mm (0.63 in)

20 Nm (14.75 lbf ft)

Torque for NPT cable glands and Conduit tubes

Ʋ Plastic housing

max. 10 Nm (7.376 lbf ft)

Ʋ Aluminium/Stainless steel housing

max. 50 Nm (36.88 lbf ft)

Input variable
Measured variable

Level of solids

Min. dielectric constant of the medium

ε

r

≥ 1.5

4

5

3

1

2

4

5

3

1

2

Fig. 39: Measuring ranges - VEGAFLEX 82
1 Reference plane

2 Probe length L

3 Measuring range

4 Upper dead band (see diagrams under Accuracy - grey section)

5 Lower dead band (see diagrams under Accuracy - grey section)

Output variable
Output

Ʋ Physical layer

Digital output signal according to standard EIA-485

Ʋ Bus specifications

Modbus Application Protocol V1.1b, Modbus over serial

line V1.02

Ʋ Protocols

Modbus RTU, Modbus ASCII, Levelmaster

Max. transmission rate

9.6 Kbit/s