VEGA VEGAFLEX 82 Modbus and Levelmaster protocol User Manual
Page 29
29
6 Set up with the display and adjustment module
VEGAFLEX 82 • Modbus and Levelmaster protocol
41833-EN-130314
Info: Instrument name, hardware and software version, calibration
date, instrument features
Note:
For optimum adjustment of the measurement, the individual submenu
items in the main menu item "Setup" should be selected one after
the other and provided with the correct parameters. If possible, go
through the items in the given sequence.
The procedure is described below.
Here you can assign a suitable measurement loop name. Push the
"OK" key to start the processing. With the "+" key you change the sign
and with the "->" key you jump to the next position.
You can enter names with max. 19 characters. The character set
comprises:
•
Capital letters from A … Z
•
Numbers from 0 … 9
•
Special characters + - / _ blanks
In this menu item you select the distance unit and the temperature
unit.
With the distance units you can choose between m, mm and ft and
with the temperature units betwenn °C, °F and K.
In this menu item you can enter the probe length or have the length
determined automatically by the sensor system.
When choosing "Yes", then the probe length will be determined
automatically. When choosing "No", you can enter the probe length
manually.
In this menu item you can select which type of medium you want to
measure. You can choose between liquid or bulk solid.
Setup - Measurement
loop name
Setup - Units
Setup - Probe length
Setup - Medium type